100 kip Capacity Load Cells

 
The load cells mentioned below are designed for 100 kip loading capacity. The designation number on each load cell is defined as follows: LC (load cell) - A (aluminum) - # (capacity, kip) - # (consecutive no.) - # (cross section area, in2) - # (inner diameter, in) - # (length, in).
 
Designation Number Outer Diameter (in)

Excitation Voltage (V)

Resistance for Balance Resistor (kΩ, between...)

Calibration Equation:

Calibration Data (VDC)
Uout= UcalA UcalB
LC-A-100-1-4.1-1.6-6.0 2.801 6.00 13 (B and D) 9.99F + 0.0303 1.01 -1.01
LC-A-100-2-4.1-1.6-6.0 2.809 6.00 13 (B and D) 9.99F + 0.0088 1.01 -1.01
LC-A-100-3-4.1-1.6-6.0 2.807 6.00 None (A and C) 9.99F-0.0132 1.03 -1.01
LC-A-100-4-4.1-1.6-6.0 2.806 6.00 - 9.99F - 0.002 1.03 -1.01
LC-A-100-5-4.1-1.6-6.0 2.799 6.00 24 (B and D) 9.99F - 0.0092 1.02 -1.01
LC-A-100-6-4.2-1.6-6.0 2.810 6.00 40 (B and D) 9.98F- 0.1863 1.01 -1.01
LC-A-100-7-4.1-1.6-6.0 2.811 6.00 24 (B and D) 9.99F - 0.1320 1.03 -1.00
LC-A-100-8-4.1-1.6-6.0 2.8 6.00 - 9.97F + 0.0132 1.05 -1.05
LC-A-100-9-4.1-1.6-6.0 2.8 6.00 8/9 (B and D) 9.98F- 0.1181 1.02 -1.00
LC-A-100-10-4.1-1.6-6.0 2.8 6.00 30 (B and D) 9.98F - 0.15013 1.05 -1.00
LC-A-100-11-4.1-1.6-6.0 2.8 6.00 25 (B and D) 9.99F - 0.0044 1.03 -1.01
LC-A-100-12-4.1-1.6-6.0 2.8 6.00 - 9.99F- 0.0134 1.03 -1.01
LC-A-100-13-4.1-1.6-6.0 2.8 6.00 30 (B and D) 9.99F - 0.0684 1.02 -1.00
LC-A-100-14-4.1-1.6-6.0 2.8 6.00 leads: metal tabs missing 0.0999F -0.0075 1.02 -1.01
LC-A-100-15-4.1-1.6-6.0 2.8 6.00 - 9.99F- 0.0138 1.02 -1.01
LC-A-100-16-4.1-1.6-6.0 2.8 6.00 40 (B and D) 9.99F - 0.0451 1.01 -1.02
LC-A-100-17-4.1-1.6-6.0 2.8 6.00 40 (B and D) 9.98F - 0.0639 1.02 -1.00
LC-A-100-18-4.1-1.6-6.0 2.8 6.00 - 9.98F- 0.1093 1.04 -1.00
LC-A-100-19-4.1-1.6-6.0 2.8 6.00 20 (B and D) 9.98F + 0.0303 1.02 -1.01