Model # |
2630-114 |
Gauge Length |
25.4 mm (2.0 in) |
Guage Length Accuracy |
± 0.5 % of Guage Length |
Maximum Travel (Axial) |
± 5.08 mm (0.2 in) |
% Travel (Axial) |
± 10 % |
Temperature Ranges |
-100 - 200°C (-148 - 392°F) |
Sensitivity (full scale) |
2.5 + 1 % - 3 % |
Repeatability |
0.1% full scale deflection |
Hysteresis |
0.3% full scale deflection |
Creep |
0.15 % full scale deflection |
Balance |
± 2.5 % full scale deflection |
Electrical calibration accuracy |
± 0.1 % full scale deflection |
Excitation |
1 to 5 Vrms, d.c. to 5 kHz |
Specimen Dimensions |
Diameter |
20 mm (0.8 in) |
Width |
0 - 40 mm (0 - 1.6 in) |
Thickness |
0 - 12.5 mm (0 - 0.5 in) |